ASTM F2193

ASTM F2193

Specifications and Testing for Spine Screws, Plates, and Rods

This standard describes the specifications and the static and fatigue bending test methods for spine screws, rods, and plates. Test set-up follows a four-point bend assembly for rods and plates. Spine screws follow a cantilever set-up. The fatigue strength of each spinal fixation component is determined by cycling to a run-out of 2.5 million cycles.

It is recommended that spine screws also be tested for torque strength, driving torque, and pullout resistance as described in ASTM F543.